Front Panel Voltage and Current Calibration

Error

Calibrate

Chapter 3 Calibration Procedures

General Calibration/Adjustment Procedure

Front Panel Voltage and Current Calibration

1Unsecure the power supply.

To calibrate the voltage and current, you must unsecure the power supply according to the procedure given on page 62.

2Disconnect all loads from the power supply and connect a DVM across output terminals.

3 Turn on the calibration mode.

Error

Calibrate

CAL MODE

Turn on the calibration mode by holding down CalibrateError (Calibrate) key as you turn on the power supply and hold down the key until you hear a long beep.

Make sure that the power supply is in ‘‘CV’’ mode. If the power supply is not in ‘‘CV’’ mode, an error occurs.

Voltage and OVP Calibration

4 Move down a level to the voltage calibration mode.

CAL SETUP 1

Error

Calibrate

The display shows the above message to indicate that the power supply is ready for voltage calibration.

5 Calibrate DAC and select the low voltage calibration point.

30 LEFT

The ‘‘START BITCAL’’ message is displayed for about 3 seconds to indicate that the power supply is ready for DAC calibration. Then it counts down numbers from 30 to 0.

V LO 0.5000 V

Then, the display shows the low voltage calibration point.

66

Page 67
Image 67
Agilent Technologies E3634A, E3633A Front Panel Voltage and Current Calibration, Cal Setup, Left, V LO 0.5000, Cal Mode