General Calibration/Adjustment Procedure

N o t e

N o t e

Chapter 3 Calibration Procedures

General Calibration/Adjustment Procedure

General Calibration/Adjustment Procedure

The calibration procedures from the front panel are described in this section. For voltage calibration, disconnect all loads from the power supply and connect a DVM across the output terminals. For current calibration, disconnect all loads from the power supply, connect an appropriate current monitoring resistor 0.01 9 across the output terminals, and connect a DVM across the terminals of the monitoring resistor.

Note that the power supply should be calibrated after 1-hourwarm-upwith no load connected.

The following table shows calibration parameters and points which should be

3

used to calibrate the output voltage and current.

 

 

Table 3-2. Parameters for Calibration

 

 

 

 

 

 

 

 

Calibration

Voltage/

Calibration Point

 

 

Parameter

Current

mnemonic

 

 

 

 

 

 

 

 

 

V LO

 

 

 

 

 

 

 

CAL SETUP 1

Voltage

V MI

 

 

 

 

 

 

 

 

 

V HI

 

 

 

 

 

 

 

CAL SETUP 2

OVP

None

 

 

 

 

 

 

 

 

 

I LO

 

 

 

 

 

 

 

CAL SETUP 3

Current

I MI

 

 

 

 

 

 

 

 

 

I HI

 

 

 

 

 

 

 

CAL SETUP 4

OCP

None

 

 

 

 

 

 

 

You can terminate any CAL SETUP without changing its calibration constants by turning off power.

Perform the voltage calibration prior to the OVP calibration and the current calibration prior to the OCP calibration.