Chapter 1 Specifications

Chapter 1 Specifications

Supplemental Characteristics

Supplemental Characteristics

Table 1-2.Supplemental Characteristics

Parameter

 

Agilent E3633A

Agilent E3634A

 

 

 

 

 

 

Output Programming Range

Low Range

0 to +8.24 V/

0 to +25.75 V/

(maximum programmable values)

 

0 to 20.6 A

0 to 7.21 A

 

 

 

 

 

 

 

High Range

0 to +20.6 V/

0 to +51.5V/

 

 

 

0 to 10.3 A

0 to 4.12 A

 

 

 

 

 

 

 

OVP

1 V to 22 V

1 V to 55 V

 

 

 

 

 

 

 

OCP

0 A to 22 A

0 A to 7.5 A

 

 

 

 

 

 

Voltage Programming

 

Full Load

No Load

Full Load

No Load

Speed: Maximum time required

 

 

 

 

 

Up

95 msec

45 msec

80 msec

100 msec

for output voltage to settle within

1% of its total excursion (for

Down

30 msec

450 msec

30 msec

450 msec

resistive load). Excludes

 

 

 

 

 

 

command processing time.

 

 

 

 

 

 

 

 

 

 

 

Remote Sensing Capability

 

 

 

 

Voltage drop

Up to 0.7 V per each lead

 

 

 

Load regulation

Add 5 mV to spec for each 1-volt change in the + output

 

lead due to load current changes.

 

 

Load voltage

Subtract voltage drop in load leads from specified output

 

voltage rating.

 

 

 

 

Temperature Coefficient, ±(% of output + offset)

Maximum change in output/readback per °C after a 30-minute warm-up

Voltage

0.01% + 3 mV

Current

0.02% + 3 mA

Stability, ±(% of output + offset)

Following 1 hour warm-up, change in output over 8 hours under constant load, line, and ambient temperature

Voltage

0.02% + 1 mV

Current

0.1% + 1 mA