V LO 0.4500 V

Error

Calibrate

Error

Calibrate

Chapter 3 Calibration Procedures

General Calibration/Adjustment Procedure

6Read the DVM and change the low voltage value on the display to match the measured voltage.

For example, if the DVM reading is 0.4500 V, adjust the voltage to 0.4500 V using the knob and resolution selection keys.

V LO 0.4500 V

Error

7Pressing Calibrate (Calibrate) key saves the change and selects the middle voltage calibration point. (‘‘V MI 25.000V’’ - E3634A model)

3

V MI 10.000 V (E3633A)

If the entered number is within an acceptable range, an ‘‘ENTERED’’ message appears for one second. If the entered number is not correct, an ‘‘MIN VALUE’’ or ‘‘MAX VALUE’’ message appears for one second and the display shows the low voltage calibration point again. The display now shows the middle voltage calibration point.

8Read the DVM and change the middle voltage value on the display to match the measured voltage.

For example, if the DVM reads 10.001 V, adjust the voltage to 10.001 V using the knob and arrow keys.

V MI 10.001 V

Error

9Pressing Calibrate (Calibrate) key saves the changes and selects the high voltage calibration point. (‘‘V HI 49.500 V’’ - E3634A model)

V HI 19.500 V (E3633A)

If the entered number is within an acceptable range, an ‘‘ENTERED’’ message appears for one second. If the entered number is not correct, an ‘‘MIN VALUE’’ or ‘‘MAX VALUE’’ message appears for one second and the display shows the middle voltage calibration point again. The display now shows the high voltage calibration point.

67

Page 68
Image 68
Agilent Technologies E3634A manual V LO 0.4500, V MI 10.000 V E3633A, V MI 10.001, V HI 19.500 V E3633A