CAL MODE

Error

Calibrate

Power

Chapter 3 Calibration Procedures

Aborting a Calibration in Progress

20Pressing the ‘‘Calibrate’’ key saves the new OCP calibration constants and return to the calibration mode.

CAL MODE

A ‘‘CALIBRATING’’ message appears for several seconds to indicate that the OCP calibration is progressing and new OCP calibration constants of ‘‘SETUP 4’’ are stored. Then the display will return to the calibration mode.

If the calibration fails, a ‘‘OCP CAL FAIL’’ message appears for one second and the display shows the ‘‘CAL SETUP 4’’ for OCP calibration again.

Turn off the power supply to exit the calibration mode.

Aborting a Calibration in Progress

Sometimes it may be necessary to abort a calibration after the procedure has already been initiated. You can abort a calibration at any time by turning the power supply off from the front panel. When performing a calibration from the remote interface, you can abort a calibration by issuing a remote interface device clear message or by pressing the front-panel ‘‘Local’’ key.

71

3

Page 72
Image 72
Agilent Technologies E3633A, E3634A manual Aborting a Calibration in Progress, Cal Mode