36 Chapter 1
Making Basic Measurements
Measuring Low Level Signals
12.To see the signal more clearly, enter 0 dB. Zero decibels of
attenuation makes the signal more visible. See Figure1- 28.
Figure 1-28 Using 0 dB Attenuation
CAUTION Before connecting other signals to the analyzer input, increase the RF
attenuation to protect the analyzer input: press Attenuation so that Auto
is underlined or press Auto Couple.
Measuring Low Level Signals Example 2:
The resolution bandwidth can be decreased to view low level signals.
1. Connect a signal generator to the analyzer input.
2. Set the signal generator frequency to 300M Hz with an amplitude of
80 dBm.
3. On the analyzer, perform a factory preset by pressing Preset,
Factory Preset (if prese nt).
4. Set the center frequency of the analyzer to 300MHz by pressing
FREQUENCY, Center Freq, 300, MHz.
5. Set the span to 5 MHz by pressing SPAN, Span, 5, MHz.
6. Set the resolution bandwidth to spectrum analyzer coupling by
pressing BW/Avg, Res BW (SA).
7. Set the Y-Axis Units to dBm by pressing AMPLITUDE, More,
Y-AxisUnits, dBm .
8. Set the reference level to 40dB m by pressing AMPLITUDE, Ref Level,
40, dBm.