Chapter 2 85
Making Complex Measurements
Making a Reflection Calibration Measurement
Reflection Calibration
1. Connect the DUT to the direction a l bridge or coupler as shown in
Figure 2-13. Terminate the unconnected port of the DUT.
NOTE If possible, use a coupler or bridge with the correct te st po rt co n n e cto r
for both calibrating and measuring . Any adapter between the test port
and DUT degrades coupler/bridge directivity and system source match.
Ideally, you should use the same adapter for the calibration and the
measurement. Be sure to terminate the second port of a two port device.
2. Connect the tracking generator output of the analyzer to the
directional bridge or c oupler.
3. Connect the analyzer input to the coupled port of the directional
bridge or coupler.
4. Perform a factory preset by pressing Preset, Factory Preset (if
present).
5. Set the Y-Axis Units to dBm by pressing AMPLITUDE, More,
Y-AxisUnits, dBm.
6. Turn on the tracking generator and if necessary, set the output
power to 10 dBm by pressing Source, Amplitude (On), 10, dBm.
CAUTION Excessive signal input may damage the DUT. Do not exceed the
maximum power that the device und er test can tolerate.
7. Set the span to 100 MHz by pressing SPAN, Span, 100, MHz.
8. Set the center frequ e n c y to 200 MHz by pressing FREQUENCY,
Center Freq, 200, MHz.
9. Set the resolution bandwidth to 3MHz by pressing BW/Avg, Res BW,
3, MHz.
10.Replace the DUT with a short circuit.
11.Normalize the trace by pressing Trace/View, More, Normalize,
StoreRef (13), Normalize (On). See Figure2-14.
This will activate the trace 1 minus trace 3 function and display the
results in trace 1. The normalized trace or flat line represents 0 dB
return loss. Normalization occurs ea ch sw eep. Replace the short
circuit with the DUT.
NOTE Since the reference trace is stored in trace 3, changing trace 3 to
Clear Write will invalidate t h e n o rm alization.