Bacharach 24-9351 manual Calibration, Sensor Check

Models: 24-9351

1 100
Download 100 pages 33.06 Kb
Page 49
Image 49
5.0 CALIBRATION

PCA

Calibration

5.0 CALIBRATION

NOTE: Bacharach recommends that the PCA be calibrated by your nearest Bacharach Service Center. Calibration, however, can be performed in the field if your facility has the necessary equipment and qualified personnel to perform the procedures described in the sections that follow.

IMPORTANT! To prevent the loss of data during calibra- tion, perform the following procedures with fresh batteries, or using an optional Power Supply (see Section 3.2).

5.1 Sensor Check

IMPORTANT! Before turning on the analyzer or perform- ing any of the calibration procedures, ensure that the ana- lyzer will be sampling fresh air, and that the probe is at room temperature.

When the analyzer is first turned on and allowed to cycle through its

60 second warmup period, and while sampling fresh air, the sensors are checked (read) and calibrated (set) to the following ambient conditions:

Oxygen sensor is spanned to 20.9%

Carbon Monoxide sensor (if installed) is zeroed

Nitric Oxide sensor (if installed) is zeroed

Pressure sensor (if installed) is zeroed

Instruction 24-9351

5-1

Page 49
Image 49
Bacharach 24-9351 manual Calibration, Sensor Check