5.2 Calibration Fixtures

Calibration

PCA

5.2 Calibration Fixtures

A gas and a draft fixture will be required to perform the various calibra- tion procedures described in this manual.

Material Required:

Calibration Kit (Refer to Section 8.2)

Calibration Gas Cylinder (Refer to Section 8.2)

Bellows

Micromanometer

Procedure:

Assemble the appropriate fixture, shown in Figure 5-1, as required by the calibration procedure being performed.

GAS

 

Leave Top

 

FIXTURE

 

Port Open

 

 

 

 

 

2

3

4

3

5

1

MENU

3

PCA

Parts Shown:

6

 

1.Gas Cylinder

2.Regulator*

DRAFT

3.

Tubing*

FIXTURE

4.

Tee*

5.

Flowmeter*

 

6.Fitting, Gas*

7.Fitting, Draft*

* Contained in Calibration Kit

MENU

PCA

7

3 4 3

Bellows

Micromanometer Material Required:

Figure 5-1. Calibration Fixtures

5-2

Instruction 24-9351