Bacharach 24-9351 manual 1.TROUBLESHOOTING GUIDE Cont, Fault, Probable Cause & Remedy

Models: 24-9351

1 100
Download 100 pages 33.06 Kb
Page 76
Image 76

Troubleshooting

PCA

TABLE 7-1. TROUBLESHOOTING GUIDE (Cont.)

 

 

Fault

Probable Cause & Remedy

 

 

NX-SENSOR ERROR code ap-

a. Calibration was attempted while

pears in the Sensor Status Screen.

sampling combustion gases.

 

b. Nitric Oxide sensor is expired.

 

Replace sensor per Section 6.7.

 

c. Bias battery on the Nitric Oxide

 

circuit board is dead. Replace bat-

 

tery per Section 6.7.2.

 

 

CO-SENSOR ERROR code ap-

a. Calibration was attempted while

pears in the Sensor Status Screen.

sampling combustion gases.

 

b. Carbon Monoxide sensor is ex-

 

pired. Replace sensor per Sec-

 

tion 6.8.

 

 

DRAFT-SENSOR ERROR code ap-

a. Sensor is exposed to pressure out-

pears in the Sensor Status Screen.

side of its detectable range.

 

b. Sensor defective. Return analyzer

 

to Bacharach for repair.

 

 

TA-SENSOR OR TL-SENSOR

a. Calibration was attempted while

ERROR code appears in the Sen-

sampling combustion gases.

sor Status Screen.

b. Thermocouple defective. Re-

 

 

place probe assembly.

 

c. Thermocouple not connected.

 

Connect thermocouple to ana-

 

lyzer per Section 3.3.

 

d. Instrument exposed to temp-

 

erature outside it’s allowable

 

operating temperature.

 

 

****” appears in one or more value

The field’s associated sensor is not

fields.

installed.

 

 

 

 

7-4

Instruction 24-9351

Page 76
Image 76
Bacharach 24-9351 manual 1.TROUBLESHOOTING GUIDE Cont, Fault, Probable Cause & Remedy, NX-SENSORERROR code ap