Bacharach 24-9351 TEST GAS CO/H2, Co-Valuexxxx Enter Co-Value, CALIBRATE H2, 1050, 1000, Measured

Models: 24-9351

1 100
Download 100 pages 33.06 Kb
Page 58
Image 58
TEST GAS CO/H2

Calibration

PCA

7.Turn off the flow of CO calibration-gas; then remove the calibration- gas cylinder from the calibration fixture.

8.Attach a CO/H2 calibration-gas cylinder to the calibration fixture; then adjust the regulator of the calibration fixture for a flowmeter reading of approximately 2 SCFH.

9.Use the st and CO-VALUE XXXXCALIBRATE H2 keys to enter a CO-VALUE that is the same as

the CO concentration which is stamped on the CO/H2 calibration-gas cylinder.

Typical Test Gas CO/H2 Screen During Calibration Procedure:

TEST GAS CO/H2

-------------------

CO-VALUE XXXX

ENTER CO-VALUE

10.Press the 1050 key to save the CO-VALUE and display the Calibrate H2 Screen.

11.After calibration gas has been applied for approximately 3 minutes (to allow for stabilization), use the st and 1000Manual background keys to enter an APPLIED

value* that equals the H2 concentration stamped on the CO/H2 calibration-gas cylinder.

Typical CALIBRATE H2 Screen During Calibration Procedure:

CALIBRATE H2

 

MEASURED

1050

PPM

APPLIED

1000

PPM

12.Press the Manual background key to calibrate the analyzer’s MEASURED value to that of the APPLIED value. At this time the Calibrate Menu Screen is redisplayed.

*The calibration range for this screen is 500–1500 ppm, any attempt to calibrate outside this range will cause the unit to display an error message.

5-10

Instruction 24-9351

Page 58
Image 58