5.4 Calibrate TA-Zero

Calibration

PCA

5.4 Calibrate TA-Zero

 

Material Required:

 

Thermocouple Simulator (K-type)

Range: 0 to 300°C

 

Accuracy: ±0.3°C

Procedure:

 

1.With the analyzer turned off, first plug the simulator's K-type connec- tor into the T-STACK jack (Figure 3-2); then turn on the analyzer and wait for its warmup cycle to complete.

2.Enter the Calibration Menu Screen per Section 5.3; then choose TA-ZERO to zero the analyzer’s stack temperature channel.

3.Adjust the simulator to 0° C (32° F).*

4.Wait until the MEASURED reading on the screen stabilizes. Then use the st and CALIBRATE TA-ZEROManual background keys to enter an APPLIED value that equals 0 °C (32 °F).

Typical Calibrate TA-Zero Screen During Calibration Procedure:

CALIBRATE TA-ZERO

MEASURED 3.0°C

APPLIED 0000.0°C

4.Press the Manual background key to calibrate the analyzer’s MEASURED value to that of the APPLIED value. At this time the Calibrate Menu Screen is redisplayed.

*The calibration range for this screen is 0 – 5 °C (32 – 41°F). Any attempt to calibrate outside this range will cause the analyzer to display an error message.