6.4 Water Trap/Filter Assembly Maintenance

PCA

Maintenance

6.4 Water Trap/Filter Assembly Maintenance

The Water Trap / Filter Assembly removes water condensate from the gas sample, and also prevents soot from contaminating the internal compo- nents of the analyzer.

Drain the water condensate after every use.

Procedure:

1.Pull off the end-cap from the Inlet End of the Water Trap / Filter Assembly (Figure 6-3).

2.Pour out all of the water condensate, and replace the end-cap.

Replace the filter element when it becomes excessively dirty.

Equipment Required:

Filter Element (Refer to Section 8.0)

Procedure:

1.Pull off the end-cap from the Outlet End of the Water Trap / Filter Assembly (Figure 6-3).

2.Remove and discard the old filter element.

3.Install a new filter element and replace the end-cap.

FilterElement RiserTube

Outlet

Inlet

End

End

Figure 6-3. Water Trap/Filter Assembly

Instruction 24-9351

6-5

Page 65