7.2 Error Codes

Troubleshooting

PCA

7.2 Error Codes

If one of the following messages appear at the bottom of the Sensor Status Screen, refer to Section 7.3 Troubleshooting Guide for information on how to correct the error.

O2 - SENSOR ERROR

O2 Sensor not connected, or is expired, or

 

was exposed to combustion gases during

 

warmup.

CO - SENSOR ERROR

Carbon Monoxide sensor is expired, or was

 

exposed to Carbon Monoxide during

 

warmup.

NX - SENSOR ERROR

Nitric Oxide sensor is expired, or was

 

exposed to Nitric Oxide during warm up,

 

or the bias battery is dead.

TL-SENSOR ERROR

Room air thermocouple is outside the

 

range of –20 to 100 °C (–4 to 212 °F)

TA-SENSOR ERROR

Flue Gas thermocouple is not connected

 

or is outside the range of –20 to 1200 °C

 

(–4 to 2192 °F)

DRAFT-SENSOR ERROR

Outside the range of –7.5 to +7.5 mb

 

(– 3 to + 3inches of water column)

LOW BATTERY

Battery voltage has dropped below 3.9 volts.

 

The instrument will shut off when battery

 

voltage drops below 3.5 volts.

“– – – –”

Not calculated (O2 above 18.8%,

 

or the temperature is above 1200 °C

 

(2192 °F), or sensor error

“* * * *”

Sensor not installed

“XXXX”

Overrange (numeric)

7-2