Bacharach 24-9351 Troubleshooting Guide, 1.TROUBLESHOOTING GUIDE, Fault, Probable Cause & Remedy

Models: 24-9351

1 100
Download 100 pages 33.06 Kb
Page 75
Image 75
7.3 Troubleshooting Guide

PCA

Troubleshooting

7.3 Troubleshooting Guide

The following table lists the causes and remedies for most of the problems that may arise with the analyzer. For help with any problem not dis- cussed here, contact your nearest Bacharach Service Center.

TABLE 7-1. TROUBLESHOOTING GUIDE

Fault

Probable Cause & Remedy

 

 

Analyzer completely nonfunctional;

a. Batteries dead. Replace batteries

won’t turn on when the

 

key is

per Section 3.2.

 

pressed.

b. Loose battery connector. Disas-

 

 

 

 

 

 

semble analyzer and ensure that

 

 

 

the battery connector is attached

 

 

 

to the printed circuit board.

 

 

 

c. Microprocessor needs to be reset.

 

 

 

Press RESET button (Fig. 4-7).

 

 

 

d. Optional Power Supply defective.

 

 

 

Replace Power Supply.

 

 

 

e. Analyzer defective. Return to

 

 

 

Bacharach for repair.

 

 

Display Screen is blank when ana-

a. Microprocessor needs to be reset.

lyzer is turned on, but pump runs

Press RESET button (Fig. 4-7).

during warmup cycle.

b. Analyzer defective. Return to

 

 

 

 

 

 

Bacharach for repair.

 

 

LOW BATTERY message appears

Batteries close to being discharged.

at bottom of display.

Analyzer will run for several min-

 

 

 

utes before the instrument shuts off.

 

 

 

Replace batteries per Section 3.2.1.

 

 

O2-SENSOR ERROR code ap-

a. Calibration was attempted while

pears in the Sensor Status Screen.

sampling combustion gases.

 

 

 

b. O2 sensor is either expired, not

 

 

 

wired correctly, or not connected to

 

 

 

the circuit board. Replace or check

 

 

 

wiring of sensor per Section 6.6.

 

 

 

 

 

 

 

 

Instruction 24-9351

7-3

Page 75
Image 75