Introduction

H25-IR

 

 

1.8 Menu System

Press the ENT and ESC keys at the same time to access the Main menu.

The menu system enables the operator to perform such functions as setting up the instrument, viewing logged readings, configuring the four user-defined setups, entering calibration data, and viewing the instrument’s diagnostic readings.

The buttons on the instrument’s front panel and probe are mapped the same and can be used inter- changeably, with exception of the CAL button.

Select a menu item by first using the Up and Down buttons to highlight the desired item, and then pressing ENT to select that item and display its data screen. After an item has been selected, use the Left, Right, Up, and Down buttons as necessary to highlight and change data within a screen. Pressing the ENT button saves the new data, while pressing the ESC button aborts the operation and displays the previous screen.

Main Menu:

 

 

 

SYSTEM

System Menu:

 

 

LOG

GAS TYPE

Gas being monitored

SETUPS

LEAK UNITS

Leak rate units of measure

CAL DATA

RELAYS

Relay set up for activating external equipment

DIAGNOSTICS

LOOP

420 mA signal’s mA per ppm loop factor

FACTORY*

BRIGHTNESS

Brightness level of display

 

CLOCK

Date and time

 

 

PCHK MODE

Determines when a pressure check is performed

 

PROBE BUTTON

Selects function of button on the Standard Probe

 

RESET DEFAULTS

Resets user programmable parameters to factory defaults

 

Log Menu:

(calibration data is not changed)

 

 

 

 

VIEW LOG

View individual logged measurements

 

CLEAR LOG

Clear total contents of log

 

LOG TOTAL

View a numerical total of all measurements stored in log (up to 50)

 

LOG LIMIT

Total accumulated leak setpoint

 

Setup Menu:

SETUP NAME

Text name given to the SETUP selection

 

SETUP NO.1

RANGE

Sensitivity of instrument (small, medium, large)

 

SETUP NO.2

GAS TYPE

Gas being monitored

 

SETUP NO.3

FEEDBACK MODE

Gas-level feedback (continuous, setpoint, H25C)

 

SETUP NO.4

SETPOINT

Gas value for the Feedback Mode & Search Auto

 

 

SEARCH AUTO ZERO ....

Sets condition when auto zero is turned OFF

 

Calibration Data Menu:

 

 

CAL TYPE

Internal or external leak source

 

EXT RATE

Leak rate of external leak source

 

EXT GAS TYPE

Type of gas in external leak source

 

CAL FACTOR

Calibration factor

 

 

RESTORE FACTORY CAL... Restores factory leak rate calibration (internal leak source only)

 

Diagnostic Menu:

 

 

 

SENSOR DATA

Raw data from all sensors

 

LAST FAULT

List of all current faults, or last fault

 

IR EMITTER

Infrared emitter’s voltage, amperage, wattage, & resistance

 

PRESSURE

All measured pressures

 

Factory Menu:

 

 

 

AUTO GAIN

Automatically set digital potentiometer (zero gas applied)

 

DIGIPOT

Manually set digital potentiometer (zero gas applied)

 

LEAK RATE

Leak rate of internal leak source

 

LEAK GAS

Gas type used in internal leak source

 

LEAK TEMP

Specified temperature for internal leak source in °C

 

LEAK TEMPCO

Temperature coefficient for internal leak source in % per °C

 

SENSOR CAL

Pressure, temperature, and flow calibration settings

 

MODEL SELECT

Selects model of unit according to part number

 

PROBE TYPE

Selects type of probe connected to instrument

 

FIRMWARE

Date firmware was compiled