Maintenance | |
|
|
11.Select File > Load Flash Image.
12.Click the browse button in the Choose Flash Image dialog box; navigate to the C:\RFU folder; and then highlight the bin file of the new firmware’s flash image file. In this example, the file H25IR 1_20.bin will be loaded.
Click Open.
13.Click OK to begin the flashing process.
14.A
15.Once flashing is complete: Exit the flash program; turn OFF the instrument; remove the serial cable; and reset the DIP switches back to their normal operation position.
16.Turn ON the instrument and observe the following message should appear:
FULL SYSTEM RESET!
SELECT MODEL FROM
FACTORY MENU AND
REINIT ALL DATA/CALS
NOTE: Flashing memory with the same firmware version will not cause the instrument to reset.
After flashing memory, the instrument will perform a complete system reset, causing all system parameters and sensor calibration data to be erased! Complete the flashing process by setting all the parameters contained in the Factory menu (refer to Section 5.2 Factory Menu).