4.2.3 Range Selection
The
Press the RANGE button until the desired range is displayed on either the instrument’s Data Display screen, or on the LCD of the Smart probe.
Instrument Display
Oz/yr
R134A
SM
Smart Probe Display
Range: SM, MED, LG
SM
SRCH (1)
NOTE: The range setting has no affect on the accuracy of the gas reading as displayed on the instrument when making a
The “SM” range is the highest sensitivity level, and is most likely used for fluorine based gases like R134a and most fluorine based refrigerant blends.
The “MED” range is used for chlorine and chlorine based gases like R12, R22, R123, etc. This position is also useful for locating large R134a or other fluorine based refrigerant leaks of 1−2 Oz/yr.
The “LG” range is used to locate very large leaks.
TIP: Blowing out the test area with shop air may enable the operator to locate a large leak more quickly.
4.2.4 Setup Selection
There are four
When first turned ON, the instrument returns to the last Setup selected. To select a different Setup, first press the SETUP button to display the Setup Selection menu. Next, use the Up and Down buttons to highlight the desired Setup, and then press ENT to make the selection and be returned to the Data Display screen. The name of the newly selected Setup will appear on the Data Display screen.
SETUP NO. 1
SETUP NO. 2
SETUP NO. 3
SETUP NO. 4
NOTE: The names listed in the Setup Selection menu are user defined, and may vary from what is shown above. The number of each setup, however, is what appears on the display of the Smart probe.
Instrument Display
Oz/yr
R134A
SETUP NO
SM SRCH
Setup Name and Number
Smart Probe Display
0.00SM S (1)