H25-IR

Specifications

 

 

7 Specifications

Gases Detected

Detects and measures, depending on instrument’s part number:

 

• CFCs, HCFCs, HFCs and Halogens

 

R11, R12, R22, R23, R113, R114, R123, R124, R134a, R227, R236fa,

 

R401A, R402A, R402A, R404A, R407A, R407C, R408A, R409A, R410A,

 

R500, R502, R503, R507, R508B, H1211, H1301, H2402, N1230, FA188

 

• R600a

 

• SF6

 

• CO2

 

• R227

Detection Method

Infrared

Measurement Units

oz/yr, g/yr, mL/s, PaM3/s, ppm

Measurement Range

0.03 to 5.00 oz/yr

 

0.85 to 142 g/yr

 

0.08 x 10–4to 100 x 10–4mL/s

 

0.08 x 10–5to 100 x 10–5PaM3/s

 

0 to 1,000 ppm

Operating Modes

Search: detects presence and location of gas leaks

 

Measure: calculates and displays leak rate

Response Time

0.5 to 1 second (dependent on probe-hose length)

Sensitivity

0.03 oz/yr (0.9 g/yr) all gases

Resolution

0.01 oz/yr (0.1 g/yr)

Accuracy:

 

Leak Rate

±1% of the accuracy value marked on tag attached to leak source

PPM

±10% of reading (can be improved to ±1% if recalibrated using a known

 

concentration of refrigerant gas)

Temperature Drift

±0.8% of reading per °C

Calibration

Internal: calibration performed using the unit’s built-in leak source

 

External: calibration performed using a customer supplied leak

 

source of a known gas type and leak rate

Relays

Four SPDT relays (rated 3 A, 240 VAC), each programmable to energize

 

under 11 different operating conditions

Audio Speaker

Audible indication of leak level

AC Power

100 to 240 VAC, 50/60 Hz

Power Consumption

15 Watts

Warm-Up Time

2 minutes (30 minutes to meet published specifications)

Operating Temperature

32 to 122 °F (0 to 50 °C)

Storage Temperature

4 to 122 °F (20 to 50 °C)

Ambient Humidity

5 to 90% RH, non-condensing

Dimensions

4 x 10.75 x 15.5 in.

 

(101.6 x 273.1 x 393.7 mm)

Weight

18 lb (8.2 kg)

Instruction 3015-4342