H25-IR

Preparing for Operation

 

 

2.5.6 Search Auto Zero

SEARCH AUTO ZERO

SETUP NO.1

<ALWAYS ON ALWAYS OFF BELOW SETPOINT BELOW 90% OF SP BELOW 75% OF SP BELOW 50% OF SP

The SEARCH AUTO ZERO function allows the operator to select one of the following conditions when the auto-zero function is disabled while in the Search mode:

NOTE: This function works in conjunction with the setpoint value as entered in Section “2.5.5 Setpoint.”

ALWAYS ON – The background gas level is always zeroed. If the probe is held over a gas leak, the instrument will automatically zero-out the reading.

ALWAYS OFF – The background gas level is never zeroed. If the probe is held over a gas leak, the instrument will not zero-out the reading, similar in operation to the Measure mode.

BELOW SETPOINT – The auto-zero function is ON when the leak-rate is below the setpoint value. If the leak rate goes above the setpoint value, then the auto-zero function turns OFF, allowing the probe to stay on the leak without the reading being zeroed-out.

For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are <2 Oz/yr. If the leak rate is 2 Oz/yr, then the auto-zero function turns OFF.

BELOW 90% OF SP – The auto-zero function is ON when the leak-rate is below 90% of the setpoint value. If the leak rate goes above 90% of the setpoint value, then the auto-zero function turns OFF, allowing the probe to stay on the leak without the reading being zeroed-out.

For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are <1.8 Oz/yr. If the leak rate is 1.8 Oz/yr, then the auto-zero function turns OFF.

BELOW 75% OF SP – The auto-zero function is ON when the leak-rate is below 75% of the setpoint value. If the leak rate goes above 75% of the setpoint value, then the auto-zero function turns OFF, allowing the probe to stay on the leak without the reading being zeroed-out.

For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are <1.5 Oz/yr. If the leak rate is 1.5 Oz/yr, then the auto-zero function turns OFF.

BELOW 50% OF SP – The auto-zero function is ON when the leak-rate is below 50% of the setpoint value. If the leak rate goes above 50% of the setpoint value, then the auto-zero function turns OFF, allowing the probe to stay on the leak without the reading being zeroed-out.

For example: If the Setpoint is 2 Oz/yr, then the auto-zero function is ON for leaks rates that are <1 Oz/yr. If the leak rate is 1 Oz/yr, then the auto-zero function turns OFF.

2.6 Calibration

The H25-IR has been calibrated at the factory, and should not require frequent calibrations when searching for gas leaks. When making measurements, however, the instrument should be calibrated on a periodic basis to ensure that it meets its published accuracy specification.

Instructions on how to calibrate the instrument are contained in Section 3 Calibration.

Instruction 3015-4342