Preparing for Operation | |
|
|
2.7 External Relay Connections
The
Locate the external relay connector on rear of unit (refer to Section 1.7.2 Rear View).
Secure the leads from the external device and its power source to the external relay connector as shown in the diagram below. This diagram shows a typical alarm device being connected to Relay 1. Repeat this procedure, as necessary, for the remaining relays.
Typical Relay 1 Wiring
-Each relay may be connected as normally open (NO), or normally closed (NC).
-The relay contacts are rated 3 A at 250 VAC.
-Power for the external alarm devices is supplied from an external power source.
-Jumper the ‘Neutral’ line of an external power source to the ‘Common’ terminals on the relay connector.
-Connect one end of the alarm device to the ‘NO’ terminal of the desired relay.
-The other end of the alarm device is connected to the other leg of the external power source. For protection, install an
18 | Instruction |