4.3 What to Do when a Fault Occurs

H25-IR

Operation

 

 

4.3 What to Do when a Fault Occurs

When the instrument’s diagnostics detects that a fault has occurred, the word “FAULT” appears in place of the Gas Type, and an audible beep occurs every 2 seconds.

Oz/yr

FAULT

SETUP SM SRCH

Fault has occurred

The cause of the fault can be determined by viewing the Diagnostic menu’s Last Fault function. To display the Diagnostics menu, first press both the ENT and ESC buttons on either the instrument’s front panel or Smart probe at the same time to display the Main menu. Next, use the Up and Down buttons to highlight “DIAGNOSTICS” and then press ENT to display the Diagnostics menu.

Main Menu:

SYSTEM LOG SETUPS CAL DATA DIAGNOSTICS Diagnostics Menu: FACTORY

Diagnostics Menu:

SENSOR DATA

LAST FAULT

IR EMITTER

PRESSURE

Select “LAST FAULT” to display the Last Fault screen. Refer to Section 5.1.2 Last Fault for a listing of the type of faults that are tracked by the diagnostics, and for information on how to possibly correct the fault.

Once the cause of the fault has been corrected, the FAULT message and the audible beep are automatically turned OFF.

4.4 What is PCHK?

The instrument maintains its accuracy under varying atmospheric conditions by performing an atmospheric pressure check either once every hour, or immediately after calibration (refer to Section Error! Reference source not found. Error! Reference source not found.). When a PCHK is performed, the unit beeps once, the pump stops, a valve opens exposing the pressure sensor to the atmosphere, and “PCHK” appears at the bottom-right of the display. During this time the instrument measures the current atmospheric pressure, and then uses this value along with the current temperature and the ideal gas law to compute the gas level. When the pressure check is complete (approximately 5 seconds), the unit beeps, the valve closes, and the pump restarts.

Oz/yr

R134A

SETUP

PCHK

Pressure check being performed

4.5 RS-232 Communication

The RS-232 connector located on the rear of the instrument (refer to Section 1.7.2 Rear View) is used to flash the instrument’s memory. Refer to Section 5.3 Flashing Memory.

Instruction 3015-4342

35