Operator Manual
Model 7003M
Oxygen measurement is affected by the following:
•Composition of the sampled gas
•Sample and calibration gas pressures
•Environment of sensor, cable and analyzer
Each must be considered when calibrating and operating the analyzer to obtain accurate measurements.
a.Gas Composition
Any substance in the sample that significantly interferes or "poisons" the sensor oxygen reading should be removed by a chemical and/or mechanical sample handling system (see Section
The sample and calibration gases must meet certain specifications (see Specifications in Preface) to ensure accurate measurement. Physical parameters that affect measurement include:
b.Sample and Cal Gas Pressures
The sample and calibration gases may be at different pressures as long as the pressure of each remains constant. Any variation in the pressure of either gas will result in an incorrect output reading of the same proportion.
Two methods of sample calibration are recommended:
1.Calibrate at the sample pressure, or
2.Calibrated at a different pressure, and enter the two pressure in the analyzer (Section
Refer to Section
c.Environmental Factors
If radio frequency interference (RFI) normally occurs in the operating environment, the sensor cable should be electrically shielded during installation.
•Pressure
•Humidity
•Temperature
•Gas composition
Also refer to Section
Rosemount Analytical Inc. A Division of Emerson Process Management |