Operator Manual
Model 7003M
NOTE
The low calibration point must be between the initial calibration point (step 1 of previous procedure) and zero, and the high calibration point must be between the initial calibration point and 20mA.
If the third point calibrated (step 8, previous procedure) is actually another low range (between the initial calibration point and CCL) it will override the original CCL but won't affect the initial calibration.
NOTE
The current output linearizing procedure uses second order polynomials to adjust the output linearity. The correction effect is maximized at the calibration point, and is smoothly diminished towards the endpoints.
d.Test Alarm Relays
1.Disconnect any alarm or ON/OFF- Control devices that are connected to the analyzer, before testing.
NOTE
When alarms are tested manually, any alarm or
2.Toggle the relays ON and OFF using the ALARM 1 and ALARM 2 keys.
3.Connect any alarm or ON/OFF- Control devices that were disconnected in step 1.
4.Press ENTER to exit the DIAGNOSTICS mode.
"Hld" flashes alternately with the last oxygen reading and the analyzer returns to HOLD mode.
5.Press ENTER again to return to RUN mode.
All operator settings and adjustments are made using the front keypad except if the power jumpers are to be changed. Refer to the following sections:
Rosemount Analytical Inc. A Division of Emerson Process Management |