Operator Manual

748223-K June 2002

Model 7003M

Error code E-0 displayed

Possible cause

Power interruption, power surge or other outside interference has occurred

Corrective Action

Press ACK. If this symptom occurs frequently, surge-protection is recommended.

Will not read low levels of oxygen

Possible cause

Sudden change in output for no apparent reason

Possible cause

Sensor subjected to sudden change in conditions, as when a shadow moves and exposes the sensor to an increase in temperature

Corrective Action

Locate sensor in an area shaded from sunlight and away from cold or warm drafts.

Ambient air leaking into system

Check/Test

Increase sample flow rate, block off flow of gas in and out of flow chamber, and not analyzer response

Corrective Action

If signal increases with test, isolate and seal leak.

If signal does not increase, replace or recharge sensor.

Resolution degrading with time

Possible cause

Sensor exposed to low temperatures.

Check/Test

Per Section 4-4b on page 4-8, check current output of sensor.

If the sensor is degrading, the signal will change in steps as the oxygen level changes instead of changing continuously.

Corrective Action

Relocate sensor or replace more frequently.

6-2 Routine Servicing and Troubleshooting

Rosemount Analytical Inc. A Division of Emerson Process Management