c. Error Messages | display. Table | ||
|
| list error codes and corrective actions. | |
| If an error condition occurs, | Pressing "ACK" or "ENTER" will | |
| error) or | acknowledge any alarm and the error | |
|
| message will disappear. | |
|
|
| |
| CODE | DEFINITION |
|
| Table |
|
|
CODE | DEFINITION |
5 volts power supply above 5.5 V | |
5 volts power supply below 4.5 V | |
12 volts power supply above 13.2 V | |
12 volts power supply below 10.8 V | |
sensor output too low for calibration | |
sensor missing | |
sample temperature is below 1°C | |
sample temperature is above 44°C | |
| Table |
|
|
CODE | DEFINITION |
HI | sample temperature is above 50°C |
LO | sample temperature is below 0°C |
Ahi | excessive sensor signal |
| Table |
|
|
CODE | DEFINITION |
COP | Computer Operating Properly error (error detected) |
iot | Invalid Op code Trap (internal computer command is invalid) |
power interruption occurred |
Table