Operator Manual
Model 7003M
| |
a. Performance |
|
Accuracy........................................ | ±0.25% oxygen |
Operating Range ........................... | 0 to 25% oxygen (endpoints set by user) |
Linearity ......................................... | ±0.25% oxygen |
| (with constant sample temperature after sensor zero offset correc- |
| tion) |
Zero Stability.................................. | ±0.25% oxygen/week at 25°C |
Span Stability................................. | ±0.25% oxygen/24 hours at 25°C |
Noise ............................................. | ±0.25% oxygen |
Response Time ............................. | 90% of reading in 20 seconds at 25°C |
Temperature Effects |
|
32 to 110°F (0 to 44°C) ......... | ±6% reading |
60 to 90°F (15 to 32°C) ......... | ±3% reading |
other 30°F (16°C) ranges ...... | +4% reading |
b. Physical |
|
Mounting |
|
Standard ................................ | Panel Mount |
Optional ................................. | Wall Mount, Pipe Mount |
Dimensions.................................... | 5.7 x 5.7 x 7.6 inches (14 x 14 x 19 cm) HxWxD |
Weight ........................................... | Approximately 3.0 pounds (1.36 kg) |
Enclosure....................................... | ABS Plastic, Black, NEMA Type 4X, IP65 |
Ambient Humidity .......................... | 95% relative humidity, |
Ambient Temperature.................... | 14 to 122°F |
Sensor Cable................................. | 2 twisted pairs, specified length up to 1000 feet (305 m) maximum |
Power Cable.................................. | 3 conductor, Type SJT, 18 AWG minimum |
Connection Conduit |
|
Power..................................... | 1/2 inch |
Sensor ................................... | 3/4 inch |
Rosemount Analytical Inc. A Division of Emerson Process Management |