Operator Manual
Model 7003M
c.Relay Contacts
Relay contacts are listed in Table
RELAY POS | TERMINAL LOCATIONS | ||
ALARM1 | ALARM2 | ||
| |||
NO (Normally open) | |||
NC (Normally closed) | |||
COM (Common) |
Relay positions are with power removed.
Table
d. Fail Safe Operation | g. Setting Alarm and |
| Relays |
The relays are setup to operated in "Fail Safe" configuration. In this configuration the relay coils are powered when the unit is NOT in an alarm condition. When an alarm condition occurs, power is removed from the relay coils.
If a power failure occurs, power will be removed from the relays and an alarm condition will be indicated to the device connected to the relays.
e.Relay Reset Upon Fatal Error
If a fatal error is detected, the unit will attempt to reset itself which will make both relays go into alarm mode. If the microprocessor is able to continue its operation, the relays will then be reset to their normal operating position.
f.Pressure Effects on Alarms
If the sample pressure changes, the partial pressure of oxygen, and thus the reading, will change and the alarm relays will not actuate at the same oxygen levels as they were set when the interment was last calibrated. See Section
Wherever "Alarm" is referred to in this section, the term
h.Setting Alarm Setpoints
This procedure can be used to:
•Verify alarm relay setpoints
•Change setpoints
•Enable alarms
•Disable alarms
Perform the following procedure for each Alarm setpoint, as required. Figure
Rosemount Analytical Inc. A Division of Emerson Process Management |