Operator Manual
Model 7003M
ROM "Checksum" test | |
| EEPROM Read/write test |
Before proceeding, the operator should be familiar with the keypad and indicators (Section
Operational checks and adjustments are not normally required, provided that the sensor,
Calibration with ambient air is recommended after
The following sections provide instructions for initial
a.Initial
1.Ensure that the proper sensor and sensor cable are connected to the analyzer.
2.Apply power to the analyzer.
The following internal tests are performed:
RAM Read/write test
3.If an error code flashes, refer to the error code lists in Table
After internal tests are passed, the software version is displayed for a few seconds, and the analyzer goes directly into RUN mode. "RUN" is briefly displayed, then current oxygen level.
Error codes
4.Monitor the oxygen output or display until the level stabilizes (typically 5 hours). The alarms and current output levels can be set immediately if desired.
5.Once the sensor has stabilized, the analyzer is operational and is ready for subsequent operations. Refer to Table
Alarm 1 | Mode: | Alarm |
| |
State: | Disabled |
| ||
| Setpoints: | On: 0.0 | Off: 0.1% | |
Alarm 2 | Mode: | Alarm |
| |
State: | Disabled |
| ||
| Setpoints: | On: 0.0 | Off: 0.1% | |
Range | 0% of scale: | 0% Oxygen | ||
100% of scale: | 25% Oxygen | |||
| ||||
Current Output Range |
| |||
Security |
| Disabled |
| |
Pressure | Calibration: | 76.0 cm Hg | ||
Process: | 76.0 cm Hg | |||
| ||||
Residual Zero Offset | 0 |
| ||
Calibration Status |
| Not Calibrated |
Table
Rosemount Analytical Inc. A Division of Emerson Process Management |