Model 7003M
Operator Manual
| • The last measured value is displayed | ||||
| Refer to Figure | ||||
| • The last output value remains at the | ||||
| below. In the HOLD mode: |
| |||
| • All instrument functions are locked | isolated current output connections | |||
| • Alarms and | ||||
|
|
|
| ||
| • All existing values are retained |
| in their last state | ||
| • The instrument stops responding to | • No keys are recognized except entry to | |||
| sensor signals |
| the diagnostic routing. | ||
|
|
|
|
| |
| KEY |
| DISPLAY | WHAT HAPPENS |
|
| DISPLAY |
| Nothing | Nothing |
|
| ACK |
| Nothing | Nothing |
|
| ALARM 1 |
| Nothing | Nothing |
|
| ALARM 2 |
| Nothing | Nothing |
|
| UP ARROW |
| Nothing | Nothing |
|
| DOWN ARROW |
| Nothing | Nothing |
|
| CAL |
| Nothing | Nothing |
|
| ENTER |
| "run" | Drops back to RUN mode |
|
| DISPLAY + ACK |
| Moves to DIAGNOSTIC mode |
|
Table
If the sensor cable must be disconnected, the instrument can be put in the HOLD mode, in which all existing signal levels and output current levels and output current levels are maintained. This may be useful to prevent disturbing an
1.Press the UP ARROW and DOWN ARROW at the same time to put the instrument on HOLD.
2.If the security is activated, enter the security identification (id) code and press ENTER to continue.
3.Hld, YES, flash briefly, then Hld flashes alternately with the last oxygen reading. The output signal will remain unchanged until the instrument returns to the RUN mode.
4.Press ENTER to return to the RUN mode, or continue with Section
Rosemount Analytical Inc. A Division of Emerson Process Management | Calibration and Adjustments |