Model 7003M
Operator Manual
SECTION 2
INSTALLATION
Carefully examine the shipping carton and contents for signs of damage. Immediately notify the shipping carrier if the carton or contents is damaged. Retain the carton and packing material until all components associated with the Model 7003M Percent Oxygen Analyzer are operational.
The sensors and analyzer unit may be shipped together or separately. The sensors are shipped assembly and charged, ready for use.
If rechargeable sensors are stored longer than six months, they may need to be recharged or rejuvenated if performance is marginal.
Disposable sensors are not rechargeable.
Refer to documentation supplied with sensors for more information.
The Model 7003M Percent Oxygen Analyzer is designed to meet NEMA Type 4X (CSA enclosure 4) requirements, provided that the cutouts in the bottom of the enclosure are fitted with approved conduit fittings or are properly sealed by the user with Hoffman plugs (PN
The Model 7003M is supplied with the Panel Mounting Kit PN 652527 as standard. Optional Wall Mount Kit PN 652539 and Pipe Mount Kit PN 652529 are available. Refer to the Outline and Mounting drawings provided in the rear of this manual.
NOTE:
Refer to Figure
Install the sensor in an area where temperature is relatively constant, taking care to avoid changes in exposure to sunlight or due to sudden drafts. A
a.Analyzer/Sensor Jumper Configuration
The Model 7003M is configured to match the type of sensor ordered with the analyzer. If the Signal Board or Sensor is repaired or replaced, verify rocker switch SW1 on the Signal Board is properly set.
Rechargeable Sensor: Close positions 1, 4, and 5
Disposable Sensor: Close positions 2, 4, and 6
Rosemount Analytical Inc. A Division of Emerson Process Management | Installation |