Operator Manual
Model 7003M
d.Interfering Gases
Gases that are reduced or oxidized in the range of 0.5 to 0.8VDC contribute to sensor current and can cause a readout error. Only a few gases have this characteristic. Common gases that should be avoided include SO2, Cl2, and
oxides of nitrogen (NOX).
If you suspect that any stream components are affecting the operation of the sensor, contact Rosemount Analytical Technical Services.
If contaminated, the rechargeable sensor must be rejuvenated and the disposable sensor must be replaced. Refer to the instructions supplied with the sensor.
a.Signal Board
The Signal Board is configured to match the type of sensor ordered with the analyzer. If the type of sensor is changed, or the Signal Board is repaired or replaced, verify the rocker switch position (SW1):
Rechargeable Sensor: Close positions 1,4, and 5
Disposable Sensor: Close positions 2,4, and 6
b.Power Supply Board
The Power Supply Board provides:
•±12V
•+5V (isolated) for other functions for that circuit
There are isolated and
c.Microprocessor Board
The Microprocessor Board is the functional center of the instrument. With the exception of the program chip, no user serviceable functions are on this assembly. If the Microprocessor Board is to be repaired or replaced, note the positions of the configuration jumpers for setting jumpers on replacement board.
An instrument timer circuit monitors instrument power for a power failure, power surge or other outside condition interfering with the microprocessor operation. When conditions return to normal, the circuit will attempt a system restart. If the microprocessor is still capable of operating, both alarm relays will be put into alarm condition, the screen will clear, and the output current will drop to 0 mA. Then the instrument will go through a normal
d.Alarm Circuits
Each alarm activation setpoint is designed to be at a different level from the deactivation setpoint:
•To prevent relay contacts from chattering if the oxygen level stays around the setpoint.
•+5V
•+24V
•To indicate to the microprocessor whether alarm is HIGH or LOW.
•To allow for a time lag between the time the alarm is activated and the time it is deactivated for ON/OFF- Control purposes.
Rosemount Analytical Inc. A Division of Emerson Process Management |