Model 7003M
Operator Manual
i.Setting Alarm ON and OFF Levels
Refer to flowchart Figure
From RUN mode, press ALARM 1.
To change the relay action from ALARM to ON/OFF Control function (or vise versa) press ALARM 1.
To disable ALARM 1 - Press ACK. "DIS" will flash briefly and return to RUN mode.
To set the ON setpoint for ALARM 1 - Use the ARROW keys, press ENTER. OFF is flashed briefly, then the alarm OFF setpoint is displayed with the right digit flashing.
To set the OFF setpoint for ALARM 1 - Use the ARROW keys, press ENTER. The instrument displays "run" and returns to RUN mode.
HIGH Alarm | ACTIVATES |
| ||
set for 8.24% O2 |
| ALARM ON |
| |
|
|
| ||
8.24% | ON | SETPOINT |
|
|
|
|
| ||
8.10% |
| OFF SETPOINT |
| |
|
|
| OFF |
|
|
| TIME |
|
|
LOW Alarm | ACTIVATES |
| ||
set for 8.10% O2 |
| ALARM ON |
| |
|
|
| ||
|
|
| OFF |
|
8.24% |
| OFF SETPOINT | DEADBAND | |
8.10% |
| SETPOINT |
| |
ON |
|
| ||
|
|
|
|
TIME
Figure
Rosemount Analytical Inc. A Division of Emerson Process Management | Operation |