Operator Manual
SECTION 4
CALIBRATION AND ADJUSTMENTS
When a new sensor is installed, adjust the zero offset current (Section
15)and calibrate the instrument with span gas
Refer to Figure
•Calibrate with span gas
•Enter sample and span gas pressures
KEY | DISPLAY | WHAT HAPPENS |
DISPLAY | Nothing | Nothing |
ACK | Nothing | Nothing |
ALARM 1 | Nothing | Nothing |
ALARM 2 | Nothing | Nothing |
UP ARROW | 10.8 →10.9 | Changes calibration factor |
DOWN ARROW | 10.9 →10.8 | Changes calibration factor |
CAL | "PPr" | Moves to pressure compensation mode |
ENTER | "run" | Drops back to RUN mode |
Table
a.Zero Calibration
Each oxygen sensor produces a constant residual electrical current in the absence of oxygen due to characteristics of the sensor. Typically this current is less that 0.1% apparent oxygen. It is different for each sensor, but will remain constant for the life of that sensor. The instrument can be zeroed by adjusting the residual current per Section
b.Span Calibration
1.Expose the sensor to ambient air1 and let the oxygen reading and sensor temperature stabilize.
1We recommend using ambient air in calibration, re- gardless of the range. Only when monitoring at 0 to
2.Press CAL to enter the CALIBRATION mode. Enter the security code if required.
3.CAL flashes and oxygen concentration is displayed with the right digit flashing. Output is held at last reading until returned to RUN mode.
4.Use the ARROW keys to set the instrument to the concentration of the calibration gas. See Table
1% range, can the accuracy of the instrument be slightly improved by using a gas of lower concentration (e.g., 1%) for calibration. If bottled air is used, verify the oxygen concentration against National Bureau of Standards certified calibration gas. The true oxygen concentration in cylinders is often different from the concentration indicated on the cylinder label.
Rosemount Analytical Inc. A Division of Emerson Process Management | Calibration and Adjustments |