Model 7003M
Operator Manual
1.Remove the rear cover from the analyzer.
2.Remove Power Supply Board.
3.Verify and/or set Jumpers E1, E2, E3, and E4 per Figure
4.Verify that fuse (F1) is correct for power.
5.
6.Replace the rear cover if no other connections are to be made. Do not overtighten the cover retaining screws to avoid stripping the threads in the plastic case.
Conduit Connections
Route the power cable through conduit and into the central opening in the bottom of the case. The inlet hole is sized for standard 1/2 inch conduit fittings.
WARNING
ENCLOSURE NOT GROUNDED
The
Connect Power Cable
Connect power wiring to the compression type connectors on the rear panel (Figure
1.Loosen the connector screw.
2.Strip insulation back approximately 1/4 inch (6.35 mm), twist stranded wires together. DO NOT tin braid wire or the connection can work loose.
3.Insert wire into connector about 1/4 inch (6.25 mm), ensuring that wires do not touch any other wire or connector.
4.Tighten screw.
115 VAC
Jumpers: E1 - E2
E3 - E4
Fuse: 1/4 Amp
115 VAC
Jumpers: E2 - E4
Fuse: 1/8 Amp
Power Supply Board
F1 | E3 E4 E2 E1 | F1 | E3 E4 E2 E1 |
1/4 AMP | 115 VAC | 1/8 AMP | 230 VAC |
Figure
Rosemount Analytical Inc. A Division of Emerson Process Management | Installation |