Appendix D

Specifications

Thermal

 

Temperature Range

-10 to 350 °C (14 to 662 °F)

Detector Type

80 x 60 thermal element focal plane array (FPA)

Accuracy

± 2 °C or 2 % (whichever is greater)

Repeatability

± 1 % or ± 1 °C (± 2 °F) whichever is greater

NETD (Thermal Sensitivity)

200 mK

Temperature Indication

0.1 °C or 0.2 °F

Optical

 

Field of View (FOV)

Rectangular. 20° Horizontal x 15° Vertical

Minimum Diameter

8.1 mm (0.32 in) at 61 cm (24 in)

Optical Resolution (D:S)

75:1 or better

Spectral Range

7.5 to 14 microns

Target Sighting

Single Laser (IEC 825/93 Class II, FDA LFR 1040.10 Class II

Instantaneous Field of View

4.4 mrad

Controls

 

Focus

61 cm (24 in) to infinity

Temperature Scale

°C or °F selectable

Palettes

Original, Gray, Rainbow (default), Ironbow, Gray Reverse

Measurement Modes

Automatic or Manual

LCD Backlight

Bright/Dim selectable

Operational

 

Adjustable Emissivity

.10 to 1.00 by 0.01

Liquid Crystal Display

70.5 mm (2.78 in) x 53.5 mm (2.1 in)

Reflected Background Temperature

-50 to 905 °C (-58 to 1661 °F)

Ambient Operating Temperature

0 to 50 °C (32 to 122 °F)

Relative Humidity

10 to 90 % noncondensing

Storage Temperature

-25 to 70 °C (-13 to 158 °F) without batteries

Storage Capacity

50 images

Electrical

 

Power

Rechargeable battery pack (included)

Battery Life

Minimum 3 hours continuous use

Battery Charge Time

2 hours in the Imager, 1 hour in the external charger (or until the green

 

LED turns on)

Data Transfer

USB interface, transfer time 25 sec for 50 images

Storage Device

Flash Memory

D-1

Page 67
Image 67