Remote Calibration

System Programming

Network Menu

Remote Calibration

The network menu also offers a Calibration option for use with devices that support network calibration.

7. Network

Statistics

Manual background Calibration

The Calibration screen contains four (4) lines of information:

Calib

 

007

 

S301M

 

 

Status:

Normal

Set Zero

 

 

Set Span

246

PPM

Line 1: Indicates the mode (Calib, meaning calibration), the (Modbus) address of the device to calibrate (001) and the type of device to calibrate (301D2)

Line 2: Indicates the status (Normal or In calib…) of the specified device

Line 3: Displays the function to perform (Set Zero)

Line 4: Displays the function to perform (Set Span) and the span gas concentration value (246 ppm)

5.On the first line, scroll to the device address and press Enter

6.Scroll through the devices to display the desired device* and press Enter to select.

7.The second line displays the device’s status

8.Scroll to select the desired function, Set Zero to set the device's zero, and press Enter to select.

9.Upon pressing Set Zero, the controller requests confirmation.

 

 

Calib 007

 

S301M

 

 

Status:

Normal

 

 

Set Zero

 

 

 

 

 

 

 

 

Set Span

246

PPM

*The device must be configured in the 96D’s database in order to be included in the device addresses displayed on screen.

M-511401

96d User Manual

61

8/10

Honeywell

 

Page 64
Image 64
Honeywell 96D user manual Remote Calibration