Parameters

Parameter Name

Parameter Type

Range of Values

 

 

 

<number>

numeric

1-99

 

 

 

Comments RF Multiplexers Model Number:

The SYSTem:CTYPe? <number> command returns

"HEWLETT-PACKARD,E1366A,0,A.01.00"(E1366A), or returns "HEWLETT-PACKARD,E1367A,0,A.01.00"(E1367A), where the "0" after "E1366A" or "E1367A" is the module serial number (always 0) and "A.01.00" is an example of the module revision code number.

Example Reading the Model Number of a Card #1 Module

SYST:CTYP? 1

! Return the model number

:CPON SYSTem:CPON <number> ALL sets the selected module (card) in a switchbox to its power-on state.

Parameters

 

Parameter Name

Parameter Type

 

Range of Values

 

 

 

 

 

 

<number>

numeric

 

1-99

 

 

 

 

Comments

RF Multiplexers Power-On State: The power-on state is all

 

channels (relays) open. Note that *RST opens all channels of all

 

modules in a switchbox, while SYSTem:CPON opens the channels

 

in only the module (card) specified in the command.

Example Setting Card #1 Module to its Power-on State

 

SYST:CPON 1

 

! Sets module #1 to power-on state