ARM
The ARM subsystem selects the number of scanning cycles (1 through 32767) for each INITiate command.
Subsystem Syntax ARM
:COUNt <number> MIN MAX
:COUNt? [MIN MAX]
:COUNt ARM:COUNt <number> MIN MAX allows scanning cycles to occur a multiple of times (1 to 32767) with one INITiate command when INITiate:CONTinuous OFF 0 is set. MIN sets 1 cycle and MAX sets 32767 cycles.
Parameters
Parameter Name | Parameter Type | Range of Values |
|
|
|
<number> | numeric | 1 32767 MIN MAX |
|
|
|
Comments ∙ Number of Scans: Use only values between 1 to 32767 for the number of scanning cycles.
∙Related Commands: ABORt, INITiate:IMMediate
∙*RST Condition: ARM:COUNt 1
Example Setting Ten Scanning Cycles
This example sets an RF multiplexer for 10 scans of channels 00 through 03. When the scan sequence completes, channel 03 is closed.
ARM:COUN 10 | ! Set 10 scans per INIT command |
SCAN (@100:103) | ! Scan channels |
INIT | ! Start scan, close channel 00 |
Chapter 5 | HP E1366A/E1367A RF Multiplexers Command Reference 41 |