[ROUTe:]

The ROUTe subsystem controls switching and scanning operations for RF multiplexer modules in a switchbox.

Subsystem Syntax [ROUTe:]

CLOSe <channel_list>

CLOSe? <channel_list>

OPEN <channel_list>

OPEN? <channel_list>

SCAN <channel_list>

:MODE NONE VOLT RES FRES :MODE?

CLOSe [ROUTe:]CLOSe <channel_list> closes the RF multiplexer channels specified by channel_list. The channel_list has the form (@ccnn) where cc = card number (01-99) and nn = channel number (00-03 and 10-13).

Parameters

Parameter Name

Parameter Type

Range of Values

 

 

 

<channel_list>

numeric

cc00-cc03, cc10-cc13

 

 

 

Comments Closing Channels: To close:

a single channel, use [ROUTe:]CLOSe (@ccnn)

multiple channels, use [ROUTe:]CLOSe (@ccnn,ccnn)

sequential channels, use [ROUTe:]CLOSe (@ccnn:ccnn)

groups of sequential channels, use

[ROUTe:]CLOSe (@ccnn:ccnn,ccnn:ccnn)

or any combination.

Closure order for multiple channels with a single command is not guaranteed.

Related Commands: [ROUTe:]OPEN, [ROUTe:]CLOSe?

*RST Condition: All channels open.

Example Closing Multiplexer Channels

This example closes channels 100 and 213 of a two-module switchbox (card numbers 01 and 02).

CLOS (@100,213)

! Close channels 100 and 213.

 

100 closes channel 00 of card #1;

 

213 closes channel 13 of card #2

48 HP E1366A/E1367A RF Multiplexers Command Reference

Chapter 5