Appendix A

HP E1366A/E1367A RF Multiplexer Specifications

Input Characteristics

AC Performance

Maximum Voltage - any center or shield to any other center, shield, or chassis:

42 V Peak

Maximum Current per Channel or Common: 1 Adc or acrms

For ZL= ZS = 50Ω (E1366A)

For ZL= ZS = 75Ω (E1367A)

<10MHz <100MHz <500MHz <1.3GHz

Insertion Loss (dB)

(40°C, 65% RH)<0.3 <0.7 <1.5 <3.0

Crosstalk(dB) (Channel-Channel*, or Channel-Common)1

Maximum Power per Channel or Common: 24 W or 24 VA

General

Channels consist of non-latching armature relays.

Typical Relay Life:

With no load: 5 x 106 operations

With max. rated load: 105 operations

Note: Relays are subject to normal wearout

based on the number of operations.

 

<-90

<-80

<-60

<-40

VSWR

<1.2

<1.25

<1.35

<1.55

* With one channel closed.

1 Specification assumes 50Ω or 75Ω termination. If all channels

unterminated, derate specification by 6dB.

Risetime: <300psec

Signal Delay: <3nsec ± 50psec

Capacitance:

Center-shield <60pF

Chassis-shield <0.15μF

Module Size/Device Type: B, register-based

DC Performance

 

Connectors Used: P1

 

 

Thermal Offset: <6μV

Number of Slots: 1

 

 

 

 

 

Typical Closed Channel Resistance:

VXIbus Interface Capability:

 

 

<1Ω

 

 

 

Slave, Interrupter, A16, D16

 

 

End of Relay Life Closed Channel

Interrupt Level: 1-7, selectable

 

 

Resistance: <3Ω

 

 

Insulation Resistance

Power Requirements:

 

 

(any terminal to any terminal):

Voltage

+5

+12

(40° C, 65% RH): >108Ω

Peak module current, IPM (A):

0.10

0.18

 

Dynamic module current, IDM (A):

0.01

0.01

Watts/Slot: 3.0