Switchbox and Scanning Voltmeter Configurations
A VXIbus instrument is a module or group of modules which perform a specified function. For the HP E1300/01 mainframe, the first module in an instrument must have a logical address which is evenly divisible by 8 (16, 24, 112), and the rest of the modules in the instrument are numbered consecutively. The instrument’s secondary address is the whole number equal to the logical address divided by 8 (for a logical address of 16, 17, 18,¼23, the secondary address is 02). For instrument definition in other mainframes, see the mainframe manual. If an instrument consists of only multiplexer modules, it is a switchbox configuration. If a multiplexer module(s) is combined with a multimeter to form an instrument, that is a scanning voltmeter configuration.
Switchbox A switchbox is a multiplexer module or group of multiplexer modules which form a single instrument. A switchbox can be connected to a multimeter which is a different VXIbus instrument, to an
10 | OUTPUT 722;"TRIG EXT;DC 10" | !Sets multimeter to external trigger |
|
| and to measure DC volts |
20 | OUTPUT 70903;"OUTP ON" | !Enables "Trig Out" port |
30OUTPUT 70903;"TRIG:SOUR BUS" !Sets switchbox to receive Bus triggers
40OUTPUT 70903;"SCAN:MODE VOLT"!Set up switchbox for voltage measurements
50OUTPUT 70903;"SCAN:PORT ABUS"!Closes the appropriate tree isolation switches while scanning, automatically makes connection to the analog bus and tree terminals.
60 OUTPUT 70903;"ARM:COUN 2" !Set for two cycles through the scan list
70OUTPUT 70903;"SCAN (@100;115)"!Selects the channel list. 100 selects first channel on card 1; 115 selects last channel on card 1
80 | OUTPUT 70903;"INIT" | !Close first channel to start |
|
| scanning cycle |
90 | FOR I = 0 TO 15 | !Start count loop |
100 | ENTER 722;A | !Enter reading into variable A |
110 | PRINT A | !Print reading in variable A |