IEEE 488.2 Common Commands

 

The following table lists the IEEE 488.2 Common (*) Commands that the

 

16-Channel FET Multiplexer Modules accept. The operation of some of

 

these commands is described in Chapter 4 of this manual. For more

 

information on Common Commands, refer to the HP 75000 Series B

 

Mainframe User’s Manual (HP Model Number E1300/E1301) or the

 

ANSI/IEEE Standard 488.2-1987.

 

 

 

 

 

 

Command

Title

Description

 

 

 

 

 

 

*IDN?

Identification

Returns identification string of the switchbox.

 

 

 

*RST

Reset

Opens all channels, and invalidates current channel list for

 

 

scanning. Sets ARM:COUN 1, TRIG:SOUR IMM, and INIT:CONT OFF.

 

 

 

*TST?

Self-Test

Power-on interrupt returns +110

 

 

Incorrect MFR ID or model code returns +103

 

 

Test pass returns 0.

 

 

 

*OPC

Operation Complete

See note below.

*OPC?

Operation Complete Query

See note below.

 

 

 

*WAI

Wait to Complete

See note below.

 

 

 

*CLS

Clear status

Clears all status registers (see STATus:OPERation[:EVENt]?).

 

 

 

*ESE

Event status enable

See note below.

*ESE?

Event status enable query

See note below.

 

 

 

*ESR?

Event status register query

See note below.

 

 

 

*SRE

Service request enable

Enables status register bits (see Chapter 4 for operation).

*SRE?

Service request enable query

See note below.

 

 

 

*STB?

Read status byte query

See note below.

 

 

 

*TRG

Trigger

Triggers the switchbox to advance the scan when scan is enabled

 

 

and trigger source is TRIGger:SOURce BUS HOLD.

 

 

 

*RCL

Recall instrument state

See note below.

 

 

 

*SAV

Store instrument state

See note below.

 

 

 

Note: These commands apply to many instruments and are not documented in detail here. See the HP 75000 Series B E1300/E1301 Mainframe User’s Manual or the ANSI/IEEE Standard 488.2-1987 for more information.

76 HP E1351A/53A 16-Channel FET Multiplexer Command Reference

Chapter 5

Page 76
Image 76
HP FET MULTIPLEXER E1351A/53A manual Ieee 488.2 Common Commands, ANSI/IEEE Standard