Direct Control Register

The Direct Control Register allows you to set the controls for individual channels. Note that the DIR bit in the Status/Control Register must be set true for this register to be enabled.

base + 0E16

15

14

13

12

11

10

9

8

7

6

5

4

3

2

1

0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Write

 

 

Undefined

 

 

X

X

X

VLD*

A_D*

B_D*

C1

C0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Read

 

 

 

 

 

 

Undefined

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

C1 - C0 These configuration bits determine the measurement mode as listed below:

C1

C0

 

0

0

Volts

0

1

2-wire ohms

1

1

4-wire ohms

1

0

Thermistor

B and A Tree Isolation

Switch Disable

(B-D*, A-D*)

Valid Channel (VLD*)

Note that to obtain the Thermistor reference temperature for Thermocouple measurements, you must set the configuration bits for Thermistor. After you have obtained the reference temperature, set the bits for Volts.

A zero in either one disables the respective tree isolation switch. Generally, both disabled together equal the [ROUTe:]SCAN:PORT NONE command.

A zero indicates that the specified channel is valid. If DIR is valid and INT EN is true, a WRITE to this register will generate an interrupt.

Executing Commands

This command lists the SCPI commands and shows how to execute them with register-based programming. For further information on register-based programming see the HP E1326B Multimeter User’s Manual .

ABORt ABORt does not have an equivalent in register-based programming. This command is executed from the front panel or over HP-IB.

ARM:COUNt The ARM:COUNt command does not have an equivalent in register-based programming. To scan for a specified number of cycles, use the interrupt generated at the end of the scan list to advance through a program loop for the set number of cycles. For each pass through the loop, reset the pointer to the beginning of the scan list. The IRQ EN in the Status/Control Register must be enabled.

DISPlay:MONitor The DISPlay:MONitor[:STATe] command does not have an equivalent in register-based programming. For high-speed operation it should be set to DISPlay:MONitor OFF.

INITiate INITiate:CONTinuous is equivalent to the CNT EN bit in the Scan Control Register.

90 HP E1351A/53A Register-Based Programming

Appendix B

Page 90
Image 90
HP FET MULTIPLEXER E1351A/53A manual Executing Commands, Direct Control Register