Direct Control Register
The Direct Control Register allows you to set the controls for individual channels. Note that the DIR bit in the Status/Control Register must be set true for this register to be enabled.
base + 0E16 | 15 | 14 | 13 | 12 | 11 | 10 | 9 | 8 | 7 | 6 | 5 | 4 | 3 | 2 | 1 | 0 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Write |
|
| Undefined |
|
| X | X | X | VLD* | A_D* | B_D* | C1 | C0 | |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
Read |
|
|
|
|
|
| Undefined |
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
C1 - C0 These configuration bits determine the measurement mode as listed below:
C1 | C0 |
|
0 | 0 | Volts |
0 | 1 | |
1 | 1 | |
1 | 0 | Thermistor |
B and A Tree Isolation
Switch Disable
Valid Channel (VLD*)
Note that to obtain the Thermistor reference temperature for Thermocouple measurements, you must set the configuration bits for Thermistor. After you have obtained the reference temperature, set the bits for Volts.
A zero in either one disables the respective tree isolation switch. Generally, both disabled together equal the [ROUTe:]SCAN:PORT NONE command.
A zero indicates that the specified channel is valid. If DIR is valid and INT EN is true, a WRITE to this register will generate an interrupt.
Executing Commands
This command lists the SCPI commands and shows how to execute them with
ABORt ABORt does not have an equivalent in
ARM:COUNt The ARM:COUNt command does not have an equivalent in
DISPlay:MONitor The DISPlay:MONitor[:STATe] command does not have an equivalent in
INITiate INITiate:CONTinuous is equivalent to the CNT EN bit in the Scan Control Register.
90 HP E1351A/53A | Appendix B |