SETTling[:TIME] [ROUTe:]SETTling[:TIME] <channel_list> sets the delay between receipt of a channel closing trigger, and the pulse when the FET channel is closed. You must specify one and only one channel for each card in the instrument which has a channel in the channel list, and the <time> will then apply to all channels on the card. The range is 1 to 32768 μsec.
Parameters
| Parameter Name | Parameter Type | Range of Values | 
| 
 | 
 | 
 | 
| time | Numeric | |
| 
 | 
 | 
 | 
| channel_list | Numeric | only one channel | 
| 
 | 
 | 
 | 
Comments ∙ Digital Bus Handshaking: This command allows a 
∙Related Commands: TRIGger:SOURce DBUS, TRIGger:SOURce IMMediate
Example Setting 16 μsec SETTling TIME
| TRIGger:SOUR DBUS | Triggering on the digital bus | 
| 
 | handshake lines | 
| SETTling:TIME  | Delay = l6μsec | 
SETTling[:TIME]? [ROUTe:]SETTling[:TIME]? [MIN MAX] <channel_list> returns the
current setting for a specified channel. MIN or MAX returns the minimum and maximum allowable settings.
Parameters
| Parameter Name | Parameter Type | Range of Values | 
| 
 | 
 | 
 | 
| MIN MAX | Numeric | MIN = 1 μsec, MAX = 32768 μsec | 
| 
 | 
 | 
 | 
| channel_list | Numeric | only one channel | 
| 
 | 
 | 
 | 
Comments ∙ Returned value is given in exponential notation.
Example Set and Query the SETTling TIME
| SETT:TIME  | Sets delay = 16μsec. | 
| SETT:TIME? | Returns +  | 
| Chapter 5 | HP E1351A/53A  | 