Multimeter
Register-Based
Programming Error
Codes
Multimeter error codes related to
Table
Error Code | Cause |
|
|
|
|
0016 | No error has occurred since the last error code was read. |
|
|
0116 | Unrecognized command opcode. |
|
|
0216 | A parameter was required but a command was received. |
|
|
0316 | A parameter is invalid or out of range for the specified command. |
|
|
0416 | Reading overrun. The data buffer is full and another measurement is taken. |
| This error also occurs when a new command (opcode) is received while the |
| multimeter is placing a reading in the data buffer. Trigger Arm is disabled. |
|
|
0516 | The command or parameter received is not allowed in the two byte reading |
| mode (10 μs aperture time). |
|
|
0C16 | Command is terminated by another command. |
|
|
0E16 | The aperture time is longer than the sample rate. |
|
|
0F16 | Input overload. A potentially damaging voltage has been applied to the |
| multimeter: |
| > +- 40V between HI and LO or HI and COMMON with the range <= 8V |
| > +- 40V between LO and COMMON on any range |
| Under this condition, the multimeter disconnects itself from the input. |
| A new range must be specified to restore operation. |
|
|
|
|