s

Appendix C

HP E1351A/53A FET Multiplexer Error Messages

Table C-1 lists the error messages associated with the multiplexer modules programmed with SCPI commands. See the HP E1406A Command Module User’s Manual for complete information on error messages.

Table C-1. 16-Channel FET Multiplexer Error Messages

No.

Title

Potential Causes

 

 

 

 

 

 

–211

Trigger Ignored

Trigger received when scan not enabled. Trigger received after scan

 

 

complete. Trigger too fast.

 

 

 

–213

INIT Ignored

Attempting to execute an INIT command when a scan is already in

 

 

progress.

 

 

 

–224

Illegal Parameter Value

Attempting to execute a command with a parameter not applicable to the

 

 

command.

 

 

 

1500

External Trigger Source Already

Assigning an external trigger source to a switchbox when the trigger

 

Allocated

source has already been assigned to another switchbox.

 

 

 

2000

Invalid Card Number

Addressing a module (card) in a switchbox that is not part of the

 

 

switchbox.

 

 

 

2001

Invalid Channel Number

Attempting to address a channel of a module in a switchbox that is not

 

 

supported by the module (for example, ch. 99 of multiplexer module).

 

 

 

2006

Command Not Supported On

Sending a command to a module (card) in a switchbox that is

 

This Card

unsupported by the module.

 

 

 

2008

Scan List Not Initialized

Executing a Scan without the INIT command.

 

 

 

2009

Too Many Channels in Channel

Attempting to address more channels than available in the switchbox.

 

List

 

 

 

 

2010

Scan Mode Not Allowed On

Attempting to execute a SCAN:MODE which is not supported.

 

Thi Card.

 

 

 

 

2011

Empty Channel List

No valid channels in <channel_list>.

 

 

 

2012

Invalid Channel Range

Invalid channel(s) specified in SCAN <channel_list> command.

 

 

Attempting to begin scanning when no valid channel list is defined.

 

 

 

2017

Incorrect ARM:COUNt

Attempting to run a downloaded scan list with ARM:COUNt set to value

 

 

other than 1.

 

 

 

2600

Function Not Supported On

Sending a command to a module (card) in a switchbox that is not

 

This Card

supported by the module or switchbox.

 

 

 

2601

Channel List Required

Sending a command requiring a channel list without the channel list.

 

 

 

 

 

 

Appendix C

HP E1351A/53A FET Multiplexer Error Messages 99

Page 99
Image 99
HP FET MULTIPLEXER E1351A/53A manual HP E1351A/53A FET Multiplexer Error Messages