Analog Input and Output Calibration Procedures

Analog Input and Output Calibration Procedures

4.Connect the UDAS to the host computer and allow at least a 15 minute warm-up period before performing the adjustments.

CAUTION:

The unit may become damaged if you use an uninsulated screwdriver and accidentally touch other components.

FIGURE C.1 Locations of Potentiometers for Adjusting Analog Circuits on page C-2 shows the locations of the potentiometers used to adjust the analog circuits on a UDAS unit.

H4

H4

0 0 0 0 0 0

TP21

R24

R25

R26

R36

R38

R39

R40

R46

R45

TP27

R58

FIGURE C.1 Locations of Potentiometers for Adjusting Analog Circuits

C-2