Analog Input and Output Calibration Procedures
.
Note: On all adjustments, there is likely a range of adjustment resulting in the proper reading. Determine this range and leave the adjustment in the center of the range.
7.Set the calibrated input to 9.5068 V.
8.Adjust Pot R46 for 9.5068 V.
9.Repeat the procedure and verify the adjustments are correct.
Note: These adjustment steps are interactive and may require several iterations.
10.Set the precision voltage source to
Note: If the value you read is more than 0.039 V different from
Adjust Offset and Gain for ± 5Volt Input Range
1.Verify the termination panel is connected to a
2.Attach a voltage calibrator to channel 0 with all other channels grounded.
3.Run a program, such as SYSCHECK, to configure the UDAS for a range of ±5 V and gain of 1.
4.Set calibrated input to 0 V, or short channel 0 to analog ground.
5.Set SYSCHECK or other program to average 50 readings. Manually or automatically take readings repeatedly.
6.Adjust Pot R39 until the software displays 0 V.
Note: On all adjustments there will be a range of adjustment that results in the proper reading. Determine this range and leave the adjustment in the center of the range.
7.Set the calibrated input to 4.5734 V.
8.Adjust Pot R40 for 4.5734 V.