Intelligent Motion Systems UDAS-1001E user manual Adjust Offset and Gain for ± 5Volt Input Range

Models: UDAS-1001E

1 65
Download 65 pages 26.36 Kb
Page 59
Image 59
Adjust Offset and Gain for ± 5Volt Input Range

Analog Input and Output Calibration Procedures

.

Note: On all adjustments, there is likely a range of adjustment resulting in the proper reading. Determine this range and leave the adjustment in the center of the range.

7.Set the calibrated input to 9.5068 V.

8.Adjust Pot R46 for 9.5068 V.

9.Repeat the procedure and verify the adjustments are correct.

Manual backgroundManual background Note: These adjustment steps are interactive and may require several iterations.

10.Set the precision voltage source to -9.95117 V and check for a reading of -9.95117.

Note: If the value you read is more than 0.039 V different from -9.95117, then the following Alternate Bipolar Calibration must be used.

Adjust Offset and Gain for ± 5Volt Input Range

1.Verify the termination panel is connected to a 50-pin connector of the UDAS unit. Begin with step 2 if the unit has built-in termination.

2.Attach a voltage calibrator to channel 0 with all other channels grounded.

3.Run a program, such as SYSCHECK, to configure the UDAS for a range of ±5 V and gain of 1.

4.Set calibrated input to 0 V, or short channel 0 to analog ground.

5.Set SYSCHECK or other program to average 50 readings. Manually or automatically take readings repeatedly.

6.Adjust Pot R39 until the software displays 0 V.

Note: On all adjustments there will be a range of adjustment that results in the proper reading. Determine this range and leave the adjustment in the center of the range.

7.Set the calibrated input to 4.5734 V.

8.Adjust Pot R40 for 4.5734 V.

C-4

Page 59
Image 59
Intelligent Motion Systems UDAS-1001E user manual Adjust Offset and Gain for ± 5Volt Input Range