![Analog Output Channel 0 Offset and Gain](/images/new-backgrounds/139691/139691123x1.webp)
Analog Output Channel 0 Offset and Gain
1.Verify the termination panel is attached to a
2.Attach a multimeter to analog output channel 0.
3.Use SYSCHECK or other software to set analog output channel 0 to
4.Adjust Pot R26 for
5.Set analog output channel 0 to 9.9951 V.
6.Adjust Pot R36 for 9.9951 V (±0.2 mV).
7.Repeat these steps to verify the results.
Analog Output Channel 1 Offset and Gain
1.Verify the termination panel is attached to a
2.Attach a multimeter to analog output channel 1.
3.Use SYSCHECK or your calibration software to set analog output channel 1 to
4.Adjust Pot R24 for
5.Set analog output channel 1 to 9.9951 V.
6.Adjust Pot R25 for 9.9951 V (±0.2 mV).
7.Repeat these steps to verify the results.